Place of Origin: | Dongguan, China |
Brand Name: | DahoMeter |
Certification: | CE,FCC |
Model Number: | DH-RUV-01 |
Minimum Order Quantity: | one set |
---|---|
Price: | USD999 to 3999 per unit |
Packaging Details: | standard exporting wooden package |
Delivery Time: | 25 working days |
Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Supply Ability: | 300 sets per month |
Product Name: | UV Aging Testing Chamber | Type: | Tower Type |
---|---|---|---|
Warranty: | 1 Year |
China Professional Supplier Plastic UV Aging Test Chamber , Rubber Aging Oven
Usage:
UV aging testing machine can simulate the sun, rain and dew to damage the material, QUV using UV lamp simulate the effect of sunlight, cold rain and dew. Being tested materials are placed in a certain environment which the light and moisture are alternating. The materials are tested for several days or weeks, you can know the damage of material at outside for several months or several years, Damage types include: bleaching, discoloration, loss of light, powder, crack, turbidity, bubble-based, embrittlement, strength decline and oxidation.
Features:
(1).The installation thickness is adjustable;Easy installation.
(2).Upward spinning door is not in the way of users to operate machine;The machine is only took very small space.
(3).By using special cooling system,the ordinary tap water is allowed to be used.
(4).Very convenient software program;Will automatically alarm if operate wrongly or machine breakdown;light intensity can be programmed.
(5).The lifetime of tube can be extended if has radiation calibrator.(more than 1600 hours)
(6).Can be divided into ordinary type,light irradiation controlling type,spraying type.
Standards:
ASTM G 153, ASTM G 154, ASTM D 4329, ASTM D 4799, ASTM D 4587, SAE, J2020, ISO 4892 All current UV aging test standards.
Specification:
Model | DH-RUV-1 |
Studio size: | 1140mm × 400mm × 380mm |
Dimensions: | 1300mm × 500mm × 1460mm |
Center distance of the lamp: | 70mm |
The distance between the specimen and the nearest parallel surface of the lamp surface: | About 50mm |
Wavelength range: | UV-A wavelength range is 315~400nm/UV-B wavelength range is 290~313nm |
Radiation intensity: | 1.0W/m2 can be adjustable |
Temperature resolution: | 0.1 ° C |
Light temperature range: | 50 ° C ~ 70 ° C / temperature tolerance is ± 3 ° C |
Condensation temperature range: | 40 ° C ~ 60 ° C / temperature tolerance is ± 3 ° C |
Blackboard thermometer measurement range: | 30~80°C/tolerance is ±1°C |
Temperature control method: | PID self-tuning temperature control mode |
Humidity range: | About 45% to 70% R.H (lighting state) / 98% or more (condensation state) |
Sink requirements: | Water depth is not more than 25mm, and there is automatic water supply controller |
Standard test piece size: | 75×150mm 48pcs |
Recommended instrument environment: | 5~35°C, 40%~85%R·H, 300mm from the wall |